This website uses cookies. By using the site you are agreeing to our Privacy Policy.

Endotoxin Detection Reagents PYROSTAR™ ES-F/Plate

  • Turbidimetric Technique
  • Endotoxin Specific
  • FDA approved

PYROSTAR™ ES-F / Plate is a specific Turbidimetric lysate reagent specifically for 96-well microplates. By using an absorbance microplate reader, simultaneous measurement of multiple samples is easily made possible. The quantitative range is wide, from 0.01 to 10 EU / mL, and it can be applied to water type tests, as well as to the testing of name drugs that require dilution.

Features

  • It is possible to specifically detect endotoxin without it being affected by (1 → 3) - β-D - Glucan in the sample.
  • US Food and Drug Administration (FDA) approved (U.S. License No.1762).
  • Regarding U.S. exports, it can also be applied to the final product testing.
  • Since it can be measured with a sample volume of 50 μL per well, even a precious sample can be measured in a small amount.
  • Stable endotoxin measurement is possible, due to the minor pH impact of the measurement sample on testing.
Multi test
Code No. Product Name Quantitative Range
(EU/mL)
Quantity
293-75401
with CSE
Limulus Amebocyte Lysate
PYROSTAR™ ES-F/Plate with CSE
0.01-10 160 tests
(2.0mLx4vials)
297-75301 Limulus Amebocyte Lysate
PYROSTAR™ ES-F/Plate without CSE
0.01-10 200 tests
(2.0mL x5vials)

standard curve information

Regression equation: log(Ta) = ー0.260424 log(C) +1.10351
Correlation coefficient: ー0.9943
Endotoxin concentration
(EU/mL)
Gelation time (min.)
0.0100 46.0
46.7
43.3
0.1000 21.2
21.5
21.2
1.000 12.0
12.1
11.9
10.000 7.4
7.4
7.5

Recovery rate of spiked endotoxins with the Turbidimetric technique

Topics

  • LAL Basic Knowledge
  • Lysate reagent selection flowchart

Contact

For inquiries please click here

CONTACT

Search LAL data sheet